The Interleukin 11 (IL11) BioAssay™ ELISA Kit is a sandwich enzyme immunoassay for in vitro quantitative measurement of IL11 in human serum, plasma tissue homogenates, cell lysates, cell culture supernates and other biological fluids.
Detection Range
15.6-1,000pg/ml
Kit Components
*026070A: Microtiter Strips, 1x96 wells, Pre-coated, ready to use *026070B: Standard, 2x1vial 026070C: Standard Diluent, 1x20ml *026070D: Detection Reagent A (green), 1x120ul *026070E: Detection Reagent B (red), 1x120ul 026070F: Assay Diluent A, 1x12ml 026070G: Assay Diluent B, 1x12ml 026070H: TMB Substrate, 1x9ml *026070J: Positive Control, xxpg/ml, 1x1vial (Lyophilized) 026070K: Stop Solution, 1x6ml 026070L: Wash Buffer, 30X, 1x20ml
Precaution
The Stop Solution (026070K) included for use in this kit is an acid solution. Wear eye, hand, face, and clothing protection when using this material.
Storage and Stability
Store *026070A, *026070B, *026070D, *026070E and *026070J at -20°C. After reconstitution of *026070J, store at 4°C and use within 5 days. Store all the other components at 4°C. Unused kit is stable for 6 months after receipt. Once kit components are opened, it is highly recommended to use remaining reagents within 1 month provided this is within the expiration date of the kit. For maximum recovery of product, centrifuge the original vials after thawing and prior to removing the cap.
Assay Procedure Summary
1. Prepare all reagents, samples and standards. 2. Add 100ul standard or sample to each well. Incubate 1 hour at 37°C. 3. Aspirate and add 100ul prepared Detection Reagent A. Incubate 1 hour at 37°C. 4. Aspirate and wash 3 times. 5. Add 100ul prepared Detection Reagent B. Incubate 30 minutes at 37°C. 6. Aspirate and wash 5 times. 7. Add 90ul TMB Substrate. Incubate 10-20 minutes at 37°C. 8. Add 50ul Stop Solution. Read at 450nm immediately.