027272
Kit Type
Sandwich ELISATests
96Sample Volume
100ulSensitivity
<14.2pg/mlDetection Method
ColorimetricDetection Range
31.2-2000pg/mlSample Matrix
Serum, plasma and other biological fluidsEU Commodity Code
38220000UN DOT Shipping
UN2796 PGIIShipping Temp
Blue IceStorage Temp
4°C/-20°COxidized Low Density Lipoprotein (OxLDL) BioAssay™ ELISA Kit (Human)
The Human Oxidized Low Density Lipoprotein (OxLDL) ELISA Kit is a sandwich enzyme immunoassay for the in vitro quantitative measurement of OxLDL in human serum, plasma and other biological fluids.
Detection Range
62.5-4000pg/ml
Precision
Intra-Assay %CV: <10% Inter-Assay %CV: <12%
Kit Components
*027272A: Microtiter Plate, 1x48 wells or 1x96 wells, Pre-coated; ready to use. *027272B: Standard, 2x1vial 027272C: Standard Diluent, 1x10ml or 1x20ml *027272D: Detection Reagent A, 1x60ul or 1x120ul *027272E: Detection Reagent B, 1x60ul or 1x120ul 027272F: Assay Diluent A, 1x6ml or 1x12ml 027272G: Assay Diluent B, 1x6ml or 1x12ml 027272H: TMB Substrate, 1x4.5ml or 1x9ml *027272J: Positive Control, 2587.1pg/ml, 1x1vial 027272K: Stop Solution, 1x3ml or 1x6ml 027272L: Wash Buffer, 30X, 1x10ml or 1x20ml
Precaution
The Stop Solution (025741K) suggested for use with this kit is an acid solution. Wear eye, hand, face, and clothing protection when using this material.
Storage and Stability
Store *027272A, *027272B, *027272D, *027272E and *027272J at -20°C. After reconstitution of *027272J, store at 4°C and use within 5 days. Store all the other components at 4°C. Unused kit is stable for 6 months after receipt. Once kit components are opened, it is highly recommended to use remaining reagents within 1 month provided this is within the expiration date of the kit. For maximum recovery of product, centrifuge the original vials after thawing and prior to removing the cap.
Notes on 027272J: Positive Control
Intended Use
The positive control is from human serum sample known to contain the target protein. It will provide a positive result in the ELISA assay, which indicates that the kit is optimized and working properly.
Characteristics
OxLDL in human serum (lyophilized powder). Note: The human based serum used in this product are screened for HIV, Hepatitis A, B, C HTLV and Syphilis.
Expected Values
783.3pg/ml
Usage
Dissolve with 150ul standard diluent and let stand for 10 minutes. If powder is partly insoluble, shake gently until fully dissolved. Mix gently before use. In the experiment, add 100ul to an appropriate well of the plate and test its concentration along with standards and your specimens. Note: Upon receipt store Positive Control vial at -20°C. After reconstitution, store Positive Control vial at 4°C and use within 5 days.
Assay Procedure Summary
1. Prepare all reagents, samples and standards. 2. Add 100ul standard or sample to each well. Incubate 2 hours at 37°C. 3. Aspirate and add 100ul prepared Detection Reagent A. Incubate 1 hour at 37°C. 4. Aspirate and wash 3 times. 5. Add 100ul prepared Detection Reagent B. Incubate 30 minutes at 37°C. 6. Aspirate and wash 5 times. 7. Add 90ul TMB Substrate. Incubate 15-25 minutes at 37°C. 8. Add 50ul Stop Solution. Read at 450nm immediately.
Important Note
This product as supplied is intended for research use only, not for use in human, therapeutic or diagnostic applications without the expressed written authorization of United States Biological.